This MRI Ventilator is pneumatically powered, single-circuit, volume-constant, time-cycled. It utilizes a high-pressure drive with regulated, high internal resistance to control pressure and is considered a non-constant pressure generator. Simultaneously, this MRI Ventilator produces a flow pattern that is constant in spite of changes in lung mechanics (inspiratory square-wave).
This MRI Ventilator is designed with internal simplicity for operations and maintenance. Quick connect features allow the operator to provide for all necessary clinical situations: controlled ventilation, continuous-flow intermittent mandatory ventilation, constant positive airway pressure, as well as being adjustable to inspiratory/expiratory ratios that are infinitely variable. A wide-range pressure relief valve allows for both the prevention of barotrauma and time-cycled, pressure-relieved ventilation.
Humidifiers and air/oxygen blenders of various types can also be used, as the operator finds necessary.
|
 |
| Specifications: | |
Dimensions: 8" x 5.5" x 5.5"
Weight: Series D 5 Ibs.
Case Material: Aluminum
All Other Materials: Aluminum, brass, and plastic
Gas Inlet Pressure Range: 30 to 150 PSI
Tidal Volume Range: 0 to 2.0 L
Breaths Per Minute: 0 to 60 BPM
Inspiratory Time Range: 0.2 to 3.0 seconds
|
Inspiratory Flow Range: 0 to 100 LPM
Pressure Relief Range: 0 to 120 CMH20
IMV/CPAP Continuous: 0 to 60 LPM
Flow Range PEEP/CPAP Range: +1 to +50 CMH20
Internal Compliance: 0 CMH20
Air/Oxygen Blenders: Any capable of producing from 30 to 60 PSI
Exhalation Valves: Oceanic-P/N 15.0 reusable
|
Caution: Federal law restricts this device to be used by or on the order of a physician.
Note: Here are just a few of the demanding work enviroments where this MRI Ventilator gets the job done.
Air/Mobile, E.R., Special Prodedures, Ambulance, In-Hospital Transfers, Anesthesia, Veterinary Medicine, Animal Labs, Special Scanning Rooms, Recovery Rooms, Disaster Control.
|
| Item Number |
Description: |
03-MAG1 |
MRI Ventilator |
03-MAG2 |
Non-Magnetic Polemount |
03-MAG3 |
IMV Kit for MRI Ventilator |
03-MAG4 |
Reusable Patient Circuit |
| 03-1560 |
Disposable Breathing Circuits (Universal I.P.P.B.), available in a box of ten. |
|