The 7500FO (fiberoptic) pulse oximeter was specifically developed for MR environments.
The sensor is easy for technicians to apply by using the simple toe or finger
wrap.
The design keeps patient safety in mind too. Radiofrequency burns
and image artifacts are virtually eliminated because of the 7500FO's fiberoptic
sensor and cable design.
Use in fixed-site or mobile MR facilities for monitoring
infant, pediatric and adult patients.
The stylish,
compact 7500FO’s simple functionality, minimum 16-hour battery life,
4-hour quick recharge, and minimum of 70 hours of memory makes it well-suited
for a variety of clinical and specialty applications. The 7500FO provides
a wide range of display indicators that provide reliable readings. The 7500FO’s
flexible alarm management options include patient security mode and a user-defined
default that offers customizable patient settings. The 7500FO’s analog
output capability allows it to interface easily with other equipment.
A Pole Mount System provides secure monitor placement. Pole sold seperately. |
|
Pulse Strength Bar Graph
Color coded bar graph assists in assessing the patient's status.
An audible tone sounds with each pulse. The pitch changes
with the patient's saturation level. |
|
Pulse Quality Indicator
Detects patient movement or signal quality changes that may
require sensor site stabilization. |
|
Sensor Fault Indicator
Indicates when a sensor has become disconnected, missaligned,
failed or is incompatible with the monitor. |
|
 |
MR OPERATIONS:
Use Only Magmedix Supplied Fiber Optic Sensors. (Sensors containing elecrical
conductors will cause patient burns). |
| The Model 7500FO must be kept outside the 200 Gauss Line of the MR
field and affixed to an immovable object. Do Not Place on MR Bed. |
 |
Please Note: The power supply is not MR Safe. |
|
| Oxygen Saturation Range |
0–100% SpO2 |
| Pulse Rate Range |
18–321 BPM |
Displays
Pulse Strength:
Alarm Indicator:
Alarm Silenced:
Numeric Displays:
Low Battery:
|
LED, Bar graph, tri-color segments
LED, bi-color
LED, amber
3-digit, 7-segment LEDs, green
LED, amber
|
Saturation Accuracy (Arms)a:
|
70–100% |
| NO MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 2 digits |
± 3 digits |
| Disposable: 6000 Series |
± 2 digits |
± 3 digits |
| MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 2 digits |
NA |
|
Pulse Rate Accuracy:
|
18–300 BPM (no motion) 40–240 BPM (motion) 40–240 BPM (low perfusion) |
| NO MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 3 digits |
± 3 digits |
| Disposable: 6000 Series |
± 3 digits |
± 3 digits |
| MOTION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 5 digits |
± 5 digits |
| LOW PERFUSION |
Adults/Pediatrics |
Neonates |
| Reusable: Finger Clip |
± 3 digits |
± 3 digits |
| Disposable: 6000 Series |
± 3 digits |
± 3 digits |
|
Memory:
|
70 hours (assuming continuous operation) |
| Temperature (Operating): |
0° to +40°C (32°F to 104°F) |
| Temperature (Storage/Transportation) |
-30° C to +50°C (-2°F to 122°F) |
Humidity (Operating): Humidity (Storage/Transportation): |
10–90% noncondensing 10–95% noncondensing |
Altitude (Operating): Hyperbaric Pressure: |
Up to 12,000 meters (40,000 feet) Up to 4 atmospheres |
| Power Requirements (Mains): |
100–240 VAC 50–60 Hz |
Internal Power:
Battery:
Operating Life (fully charged):
Storage Life:
Recharge Rate:
|
7.2 volt NiMH battery pack
16 hours minimum
21 days minimum
4 hours maximum
|
| Dimensions: |
Approximately 219 mm (8.6”) W x 92 mm (3.6”) H x 142 mm (5.6”) D |
| Weight: |
Approximately 900 grams (2 lbs) with battery |
| Warranty: |
3 years |
| Classification per IEC 60601-1/CSA601.1/UL60601-1: |
| Type of Protection: |
Internally powered (on battery power) |
| Degree of Protection: |
Type BF-Applied Part |
| Mode of Operation: |
Continuous |
| Enclosure Degree of Ingress Protection: |
IPX2 |
Analog Outputs:
|
|
| SpO2 Output Range: |
0-1 VDC (0–100% SpO2), 1.27 VDC (out of track) |
| Pulse Rate Output Range: |
0-1 VDC (0–300 BPM), 1.27 VDC (out of track) |
| Event Marker: |
0 V (no event), 1 V (event occurred) |
| Accuracy: |
± 2% (SpO2), ± 5% (Pulse Rate) |
| Load Current: |
2 mA maximum |
a ± Arms represents approximately 68% of measurements
Specifications are subject to change without notice. |